External SCSI-2 Differential Adapter Bus PTC Isolation Procedure

Isolate the external SCSI bus PTC fault with the following procedure:

Notes:
  1. Only the probe tips are touching the solder joints. Do not allow the probes to touch any other part of the component.
  2. The external bus is differential.
  1. Ensure that system power and all externally attached device power is turned off.
  2. Check to ensure all devices are marked SCSI Differential and that the terminator on the end of the SCSI bus is also marked differential. If not, you may have a single-ended SCSI device or terminator on the differential SCSI bus. Single-ended devices do not work on a differential SCSI bus and may cause a PTC type error to be reported. The entire SCSI bus may appear to be intermittent. After ensuring the system is completely differential, continue.
  3. Disconnect the external cables from the adapter and remove the adapter from the system.
  4. Verify with a digital Ohmmeter that the internal PTC resistor, labeled Z1, (refer to the illustration on page External SCSI-2 Differential Adapter Bus PTC Isolation Procedure) is cool and in a low resistance state, typically less than 1/2 Ohm. Measuring across, be sure to probe both sides of the PTC resistor where the solder joints and board come together. The polarity of the test leads is not important. If necessary, allow the PTC resistor to cool and measure again.
    a4aa0149
  5. This step determines if there is a short on the adapter. Locate capacitor C1 and measure the resistance across it using the following procedure:
    1. Connect the negative lead to the side of the capacitor marked "GND". Be sure to probe at the solder joint where the capacitor and board come together.
    2. Connect the positive lead to the side of the capacitor marked "Cathode D1" on the board near C1. Be sure to probe at the solder joint where the capacitor and board come together.
      • If there is no fault present, then the resistance reading is 25 to 35 Ohms. The adapter is not faulty. Continue to the next step.
      • If the resistance measured is higher than 35 Ohms, check to see if RN1, RN2, and RN3 are plugged into their sockets. If these sockets are empty, you are working with a Multi-Initiators or High-Availability system. With these sockets empty, a resistive reading across C1 cannot be verified other than it measures a high resistance (not a short). If the resistance measurement is not low enough to be suspected as a fault (lower than 10 Ohms), continue to the next step.
      • If the resistance is high and there is no external cable to reattach, there is no apparent cause for the PTC diagnostic failure.
      • If the resistance reading is low, typically below 10 Ohms, there is a fault. Because there are no cables attached, the fault is on the adapter. Replace the adapter.
      • If the resistance measured was high and there is an external cable to reattach, proceed to the next step.
  6. Reattach the external cable to the adapter.
    1. Measure across C1 as previously described.
    2. If the resistance is between 10 to 20 Ohms, there is no apparent cause for a PTC resistor failure.
    3. If the resistance is less than 10 Ohms, there is a possibility of a fault on the external SCSI bus. Troubleshoot the external SCSI bus by disconnecting devices and terminators. Measure across C1 to determine if the fault has been removed.